Re-using residual drug during intravenous infusion treatment
To minimize drug loss residual drug from an intravenous infusion are sometimes used on multiple patients. The non-return valves in the infusion system should in theory prevent the rest of the system from being contaminated. There are currently no Swedish guidelines for if this is appropriate or not.
Reading time approx. 5 minutes
Published:
Publication type:
Question
What scientific studies are there on the use of residual drug from an intravenous infusion for more than one patient?
Identified literature
The primary studies have not been assessed for risk of bias by SBU enquiry service. | |||
Author Year Ref |
Study design | Setting | Outcome |
Azevedo et al 2020 [1] |
Experimental, in vitro | Radiology, contrast fluid | Bacterial migration |
Goebel et al 2019 [2] |
Experimental, in vivo | Radiology, contrast fluid | Bacterial migration |
Nandy et al 2017 [3] |
Experimental, in vitro | Component testing | Bacterial migration |
Cona et al 2012 [4] |
Experimental, animal | Radiology, contrast fluid | Radiotracer migration |
Ellger et al 2011 [5] |
Experimental, in vitro | Intensive care unit | Bacterial migration |
Sikora et al 2010 [6] |
Mathematical modelling | Radiology, contrast fluid | Blood-born transmission |
Radke et al 2010 [7] |
Experimental, in vitro | Anasthesia, propofol | Bacterial migration |
Buerke et al 2008 [8] |
Experimental, in vitro | Radiology, contrast fluid | Bacterial migration |
Eichler 2004 [9] |
Experimental, in vitro | Anasthesia, propofol | Bacterial migration |
Garcia et al 1996 [10] |
Experminental, in vivo | Radiology, contrast fluid | Bacterial migration |
Turner et al 1991 [11] |
Experimental, in vitro | Pain, opioid analgesia | Retrograde flow |
Trépanier et al 1990 [12] |
Experimental, human | Anasthesia | Blood detection |
References
- Azevedo MPF, Monteiro RM, Castelani C, Bim FL, Bim LL, Macedo AP, et al. Biosafety of Non-Return Valves for Infusion Systems in Radiology. Sci Rep 2020;10:9574.
- Goebel J, Steinmann J, Heintschel von Heinegg E, Hestermann T, Nassenstein K. Bacterial contamination of automated MRI contrast injectors in clinical routine. GMS Hyg Infect Control 2019;14:Doc05-Doc05.
- Nandy P, Young M, Haugen SP, Katzenmeyer-Pleuss K, Gordon EA, Retta SM, et al. Evaluation of one-way valves used in medical devices for prevention of cross-contamination. Am J Infect Control 2017;45:793-798.
- Cona MM, Bauwens M, Zheng Y, Coudyzer W, Li J, Feng Y, et al. Study on the microbial safety of an infusion set for contrast-enhanced imaging. Invest Radiol 2012;47:247-51.
- Ellger B, Kiski D, Diem E, van den Heuvel I, Freise H, Van Aken H, et al. Non-return valves do not prevent backflow and bacterial contamination of intravenous infusions. J Hosp Infect 2011;78:31-35.
- Sikora C, Chandran AU, Joffe AM, Johnson D, Johnson M. Population risk of syringe reuse: estimating the probability of transmitting bloodborne disease. Infect Control Hosp Epidemiol 2010;31:748-54.
- Radke OC, Werth K, Borg-Von-Zepelin M, Saur P, Apfel CC. Two serial check valves can prevent cross-contamination through intravenous tubing during total intravenous anesthesia. Anesth Analg 2010;111:925-928.
- Buerke B, Mellmann A, Stehling C, Wessling J, Heindel W, Juergens KU. Microbiologic contamination of automatic injectors at MDCT: experimental and clinical investigations. AJR Am J Roentgenol 2008;191:W283-7.
- Eichler W, Schumacher J, Ohgke H, Klotz KF. Reuse of a set for total intravenous anaesthesia: safe against bacterial contamination? Eur J Anaesthesiol 2004;21:501-3.
- Garcia M. Preliminary report: biosafety analysis of one-way backflow valves for multiple patient use of low osmolar intravenous contrast solution. Can Commun Dis Rep 1996;22:28-31.
- Turner GA, Murphy DF. Antireflux valves in intravenous opioid analgesia: are they necessary? Anaesthesia 1991;46:1064-5.
- Trépanier CA, Lessard MR, Brochu JG, Denault PH. Risk of cross-infection related to the multiple use of disposable syringes. Can J Anaesth 1990;37:156-9.
Literature search
PubMed via NLM 2021-04-06
Re-using residual drug during intravenous infusion treatment.
The search result, usually found at the end of the documentation, forms the list of abstracts. [MeSH] = Term from the Medline controlled vocabulary, including terms found below this term in the MeSH hierarchy; [MeSH:NoExp] = Does not include terms found below this term in the MeSH hierarchy; [MAJR] = MeSH Major Topic; [TIAB] = Title or abstract; [TI] = Title; [AU] = Author; [TW] = Text Word; Systematic[SB] = Filter for retrieving systematic reviews; * = Truncation; EED = Economic Evaluations; HTA = Health Technology Assessments |
||
Search terms | Items found | |
Intervention: | ||
1. | “Anesthesia, Intravenous” [mesh] OR “Administration, Intravenous” [mesh] OR Infusion Pumps [mesh] OR (intravenous [TIAB] OR “infusion pump*” [TIAB] OR “infusion system*” [TIAB]) | 390 520 |
2. | ((“Equipment Reuse” [mesh] OR “Multi dose” [TIAB] OR multidose [TIAB] OR “single use” [TIAB] OR singleuse [TIAB] OR “single dose” [TIAB] OR singledose [TIAB] OR “multi use” [TIAB] OR “redose” [TIAB] OR “re dose” [TIAB] OR “re use” [TIAB] OR “reuse” [TIAB] OR “multiple patient*” [TIAB]) AND (vial* [TIAB] OR ampoule* [TIAB] OR needle* [TIAB] OR inject* [TIAB] OR equip* [TIAB] OR syringe* [TIAB] OR infus* [TIAB])) | 15 193 |
3. | “Drug contamination” [mesh] OR Equipment Contamination OR “cross infection” [mesh] OR contaminat* [TIAB] OR “cross infect*” [TIAB] OR crossinfect* [tiab] OR “drug waste*” [TIAB] OR “patient to patient” [TIAB] | 278 602 |
4. | “check valve*” [TIAB] OR (("Anti reflux" OR "non return" OR "one way" OR antireflux OR nonreturn OR “back check”) AND valve* [TIAB]) | 1 351 |
5. | 2 AND 3 | 704 |
Final | 1 AND (4 OR 5) | 142 |
Embase via embase.com 2021-04-06
Re-using residual drug during intravenous infusion treatment.
/de = Term from the EMTREE controlled vocabulary; /exp = Includes terms found below this term in the EMTREE hierarchy;/mj = Major Topic; :ab = Abstract; :au = Author; :ti = Article Title; :ti,ab = Title or abstract; * = Truncation; ’ ’ = Citation Marks; searches for an exact phrase | ||
Search terms | Items found | |
Intervention: | ||
1. | 'intravenous drug administration'/exp OR 'intravenous anesthesia'/exp OR 'intravenous anesthetic agent'/exp OR 'infusion pump'/exp OR 'infusion system'/exp OR (intravenous OR “infusion pump*” OR “infusion system*”):ti,ab,kw | 716 286 |
2. | ((“Multi dose” OR multidose OR “single use” OR singleuse OR “single dose” OR singledose OR “multi use” OR redose OR “re dose” OR “re use” OR reuse OR “multiple patient*”) AND (vial* OR ampule* OR needle* OR inject* OR equip* OR syringe* OR infus*)):ti,ab,kw | 22 322 |
3. | 'drug contamination'/exp OR 'cross contamination'/exp OR 'cross infection'/exp OR 'medical device contamination'/exp OR (contaminat* OR 'cross infection' OR crossinfection OR “drug waste” OR “patient to patient”):ti,ab,kw | 279 500 |
4. | ((Check OR “Anti reflux” OR “non return” OR “one way” OR oneway OR antireflux OR nonreturn OR “back check”) NEAR/3 valve) | 726 |
5. | 2 AND 3 | 1 160 |
Final | 1 AND (4 OR 5) | 158 |